The Myron L 750 Series II Conductivity Monitor is a reliable, high-performance instrument designed for continuous monitoring of conductivity in a wide range of industrial and water treatment applications. Engineered with accuracy and durability in mind, it is compatible with CS51-series sensors, which include automatic temperature compensation and modular design for ease of maintenance. This system is ideal for users requiring dependable conductivity readings with long-term stability under demanding conditions, including high temperature and pressure.

750 Series II Conductivity Monitor Myron L

The Myron L 750 Series II Conductivity Monitor is a reliable, high-performance instrument designed for continuous monitoring of conductivity in a wide range of industrial and water treatment applications. Engineered with accuracy and durability in mind, it is compatible with CS51-series sensors, which include automatic temperature compensation and modular design for ease of maintenance. This system is ideal for users requiring dependable conductivity readings with long-term stability under demanding conditions, including high temperature and pressure.

Features

  1. Automatic Temperature Compensation

    • Integral thermistors provide real-time temperature correction to 25°C, ensuring accurate readings from 0–100°C (32–212°F).

  2. Durable Sensor Construction

    • The CS51-series sensors feature a modular polypropylene bushing and a robust metal sensing element, making them suitable for harsh environments.

  3. Versatile Sensor Options

    • Choose from CS51LC (0.1 constant), CS51 (1.0 constant), and CS52 (10.0 constant) sensors to match different conductivity ranges.

  4. Wide Environmental Tolerance

    • CS51 sensors withstand up to 100 psi at 100°C (212°F), with optional models available for pressures up to 250 psi at 205°C.

  5. Flexible Mounting and Cabling

    • Standard ¾” NPT threaded polypropylene bushings with optional ½” PVDF or 316 stainless steel fittings. Shielded cables available in 10 ft (3 m) standard lengths, up to 100 ft (30 m).

  6. Easy Maintenance and Integration

    • Modular sensor design allows low-cost replacements and simple installation, ideal for continuous process monitoring or system integration.

Applications

  • Industrial process water monitoring

  • Reverse osmosis system control

  • Boiler and cooling tower water management

  • Wastewater treatment plants

  • Laboratory conductivity analysis

  • Food and beverage water quality monitoring

Specification

  • Temperature Compensation:
    Integral thermistor, automatic to 25°C across 0–100°C (32–212°F)

  • Pressure / Temperature Limits:
    CS51: 100 psi @ 100°C (212°F), 250 psi @ 205°C (optional models CS40, CS41 available on request)

  • Sensor Types and Ranges:

    • CS51LC (K=0.1): for conductivity readings ≤ 5 µS/cm

    • CS51 (K=1.0): for 0–10 to 0–20,000 µS/cm

    • CS52 (K=10.0): for readings above 20,000 µS/cm

  • Bushing Materials:
    Modular polypropylene, threaded ¾” NPT (optional: ½” PVDF or 316 stainless steel)

  • Cable Lengths:
    Shielded cable, 10 ft (3 m) standard; 25 ft (8 m) and 100 ft (30 m) available (max recommended length: 100 ft/30 m)

  • Sensor Dimensions:
    CS51: 30 mm length x 13 mm diameter (1.2 in x 0.5 in) metal portion

  • Display and Interface:
    Paired with 750 Series II monitor/controller for real-time digital readout and alarm functionality (monitor sold separately)

Integration & Expansion

  • Compatible with Myron L 750 Series II monitoring systems

  • Ideal for long-term, high-accuracy industrial installations

  • Sensor selection based on application-specific conductivity range and environmental conditions

The 750 Series II Conductivity Monitor with CS51-series sensors delivers precision, flexibility, and rugged performance for reliable conductivity measurements across a wide range of critical water quality applications.