ACZET Moisture Analyzer MB 40 Maximum Capacity 40 gm
ACZET Moisture Analyzer MB 40 Maximum Capacity 40 gm
Description
-
The ACZET Moisture Analyzer MB 40 is a reliable, high‑performance instrument for precise moisture content determination in solid, liquid, and paste samples across industrial and laboratory applications.
-
Features a maximum capacity of 40 g with mass readability of 0.001 g and moisture readability of 0.001%, ensuring exceptional accuracy.
-
Powered by an infrared halogen heating source with a maximum temperature of 175 °C (optional 250 °C), providing fast and uniform drying.
-
Offers five heating profiles: Standard, Rapid, Step, Gentle, and High Temperature, adaptable to various sample types.
-
Equipped with multiple switch‑off criteria: automatic, manual timed, user‑defined by weight loss (G/T or %U/T), and intelligent shut‑off.
-
Includes intelligent auto method development to optimize drying parameters automatically.
-
Stores up to 40 methods (alpha‑numeric named) and 100 batches, with statistics per method and per batch for the last 9,999 results.
-
Retains the last 9,999 moisture determination results in data storage.
-
Features a graphical display with LED backlight showing % moisture, % dry substance, diagram graph, ratio, % atro moisture, % atro dry, and % gm/ltr.
-
Supports GLP/GMP compliant printing with user‑configurable print formats.
-
Provides quality management with password protection for method, setting, print format, data deletion, calibration, and drying functions.
-
Includes weighing and heater tests for routine verification.
-
Connectivity options: RS232C bidirectional, USB Ethernet, and PS2 interfaces.
-
Standby temperature adjustable from 30 °C to 100 °C.
-
User interface available in English.
Specifications
-
Maximum Capacity: 40 g
-
Precision of Readout (Mass): 0.001 g
-
Precision of Readout (Moisture): 0.001%
-
Repeatability (sd):
-
For initial sample weight of 2 g: 0.1%
-
For initial sample weight of 10 g: 0.03%
-
-
Display Information: % moisture, % dry substance, diagram graph, ratio, % atro moisture, % atro dry, % gm/ltr
-
Temperature Range: Maximum 175 °C (optional 250 °C)
-
Maximum Time Selection: 999.9 min
-
Heating Source: Infrared halogen
-
Temperature Increment: 1 °C interval
-
Heating Profiles: Standard, Rapid, Step, Gentle, High Temperature
-
Switch‑Off Criteria: Automatic, manual timed, user DEF G/T, user DEF %U/T, intelligent
-
Menu Language: English
-
Pan Size: 95 mm Ø × 8 mm
-
Statistics:
-
Method‑wise for each individual 40 methods
-
Batch‑wise for each individual 100 batches for the last 9,999 results
-
-
Data Storage: Last 9,999 moisture determination results
-
Auto Method Development: Intelligent
-
Method Memory: 40
-
Batch Memory: 100
-
Method Designation: Alpha‑numeric
-
Serial Number: Yes
-
Compile (Method Linking): Yes
-
Standby Temperature: 30 °C to 100 °C
-
Display: Graphical display with LED backlight
-
Data Connection Interfaces: RS232C bidirectional, USB Ethernet, PS2
-
Print: GLP/GMP compliant, user configurable
-
Test Functions: Weighing test and heater test
-
Quality Management:
-
Weight‑in assist: Yes
-
Password protection for: method, setting, print format, data deletion, calibration, drying
-
-
Power Supply: 230 V / 50 Hz or 115 V / 60 Hz
-
Power Consumption: Max 415 W
-
Housing Dimensions (W × D × H): 211 × 342 × 178 mm
-
Weight: Approximately 5 kg
-
Packing Dimensions (W × D × H): 325 × 390 × 285 mm